Analytical Instrumentation Questions and Answers – Ion Analyser

This set of Analytical Instrumentation Multiple Choice Questions & Answers (MCQs) focuses on “Ion Analyser”.

1. If a solid or liquid membrane is placed in pure water, the membrane dissolves slightly, producing an equilibrium concentration of the measured ion. This is represented as __________
a) A constant, Eo
b) Activity
c) Blank correction, Cb
d) Concentration of standard solution, Cs
View Answer

Answer: c
Explanation: If a solid or liquid membrane is placed in pure water, the membrane dissolves slightly, producing an equilibrium concentration of the measured ion. This is represented as a blank correction, Cb. This is the constant background for all measurements.

2. During which of the following conditions is the blank correction not necessary?
a) If sample concentration rises in linear response region
b) If sample concentration falls in linear response region
c) If sample concentration rises in non-linear response region
d) If sample concentration falls in non-linear response region
View Answer

Answer: b
Explanation: The blank correction is not necessary if sample concentration falls in the linear response region. This is the constant background for all measurements.

3. During which of the following conditions is the blank correction necessary?
a) If sample concentration rises in linear response region
b) If sample concentration falls in linear response region
c) If sample concentration rises in non-linear response region
d) If sample concentration falls in non-linear response region
View Answer

Answer: d
Explanation: The blank correction is necessary if sample concentration falls in non-linear response region. This is the constant background for all measurements.
advertisement
advertisement

4. Ion analysers need calibration once in which of the following durations?
a) Every two or three hours
b) Every nine or ten hours
c) Every 24 hours
d) Every 48 hours
View Answer

Answer: a
Explanation: Ion analysers need calibration every two or three hours. Temperature affects the output.

5. In ion analysers, which of the following serve the dual purpose of controlling the operating current of FET and providing current gain?
a) FET itself does it
b) SCR
c) Op-amp
d) Diode
View Answer

Answer: c
Explanation: In ion analysers, op-amp serves the dual purpose of controlling the operating current of FET and providing current gain. Two FETs are operated as source followers.
Sanfoundry Certification Contest of the Month is Live. 100+ Subjects. Participate Now!

6. The presence of dirt or moisture degrades which of the following?
a) Circuit components
b) High input impedance of buffer amplifier
c) Low input impedance of buffer amplifier
d) Low output impedance of buffer amplifier
View Answer

Answer: b
Explanation: The presence of dirt or moisture degrades the high input impedance of the buffer amplifier. Solder flux may also degrade it.

7. Which of the following is not a type of ground in ion analyser?
a) Chassis and electrostatic ground
b) Digital ground
c) Analog ground
d) Reference ground
View Answer

Answer: d
Explanation: There is no such point as a reference ground. Three types of grounding are provided which are chassis and electrostatic ground, digital ground and analog ground.
advertisement

8. The timing for a microprocessor and for all signals is generated by the CPU clock.
a) True
b) False
View Answer

Answer: a
Explanation: The timing for a microprocessor and for all signals is generated by the CPU clock. A crystal may be present to provide clock signals.

9. Which of the following is the only operator entry required in PC-based ion analysers?
a) Hardware set-up
b) Signal measurement
c) Calibration
d) Sample calculation
View Answer

Answer: c
Explanation: Calibration is the only operator entry required in PC-based ion analysers. The operator has to enter the concentration of calibration standards.
advertisement

10. The analog and digital grounds are connected together in ion analysers.
a) True
b) False
View Answer

Answer: b
Explanation: The analog and digital grounds are not connected together in ion analysers. This is to ensure that the digital signals never flow through the same conductor as analog signals.

Sanfoundry Global Education & Learning Series – Analytical Instrumentation.

To practice all areas of Analytical Instrumentation, here is complete set of 1000+ Multiple Choice Questions and Answers.

advertisement
advertisement
Subscribe to our Newsletters (Subject-wise). Participate in the Sanfoundry Certification contest to get free Certificate of Merit. Join our social networks below and stay updated with latest contests, videos, internships and jobs!

Youtube | Telegram | LinkedIn | Instagram | Facebook | Twitter | Pinterest
Manish Bhojasia - Founder & CTO at Sanfoundry
Manish Bhojasia, a technology veteran with 20+ years @ Cisco & Wipro, is Founder and CTO at Sanfoundry. He lives in Bangalore, and focuses on development of Linux Kernel, SAN Technologies, Advanced C, Data Structures & Alogrithms. Stay connected with him at LinkedIn.

Subscribe to his free Masterclasses at Youtube & discussions at Telegram SanfoundryClasses.