Analytical Instrumentation Questions and Answers – Estimation of Carbon Monoxide

This set of Analytical Instrumentation Multiple Choice Questions & Answers (MCQs) focuses on “Estimation of Carbon Monoxide”.

1. Which of the following is the full scale range of Infrared spectroscopy method used for measurement of carbon monoxide?
a) 0-2 ppm
b) 0-500 ppm
c) 0-1 ppm
d) 0-50 ppm
View Answer

Answer: d
Explanation: The full scale range of Infrared spectroscopy method used for measurement of carbon monoxide is 0-50 ppm. It depends on the characteristic absorption of the CO molecule at a particular wavelength.

2. Which of the following is the full scale range of Gas chromatography method used for measurement of carbon monoxide?
a) 0-200 ppm
b) 0-500 ppm
c) 0-1 ppm
d) 0-50 ppm
View Answer

Answer: a
Explanation: The full scale range of Gas chromatography method used for measurement of carbon monoxide is 0-200 ppm. The detector used in this method is a flame ionization detector.

3. Which of the following are not the characteristics of Non-dispersive infrared analyzer method?
a) Easy to maintain
b) High response speed
c) Effect of interfering components is high
d) High sensitivity
View Answer

Answer: c
Explanation: Effect of interfering components is small for Non-dispersive infrared analyzer method. It depends on the characteristic absorption of the CO molecule at a particular wavelength.
advertisement
advertisement

4. Non-dispersive infrared analyzer depends on the characteristic absorption of the CO molecule at a wavelength of __________
a) 5.4 µm
b) 4.6 µm
c) 8.9 µm
d) 10.8 µm
View Answer

Answer: b
Explanation: Non-dispersive infrared analyzer depends on the characteristic absorption of the CO molecule at a wavelength of 4.6 µm.

5. The zero of the scale is calibrated using which of the following as the sample?
a) Air
b) Water
c) Inert gas
d) Pure nitrogen
View Answer

Answer: d
Explanation: The zero of the scale is calibrated using pure nitrogen as the sample. Optical filters are used to minimize the effects of interfering gases.
Sanfoundry Certification Contest of the Month is Live. 100+ Subjects. Participate Now!

6. The disadvantage of Non-dispersive infrared analyzer is that the effect of flow rate is large.
a) True
b) False
View Answer

Answer: b
Explanation: In Non-dispersive infrared analyzer, the effect of flow rate is small. Optical filters are used to minimize the effects of interfering gases.

7. A stripper column used in gas chromatography, can be used to retain which of the following when an air sample is passed through it?
a) CO
b) Methane
c) Light hydrocarbons
d) Heavy hydrocarbons
View Answer

Answer: d
Explanation: A stripper column can be used to retain heavy hydrocarbons when an air sample is passed through it. It passes CO and methane into the column.
advertisement

8. Which of the following is not a component of the mixture present in the reference cell of Non-dispersive infrared analyzer?
a) CO
b) Nitrogen
c) Water vapour
d) Hydrogen
View Answer

Answer: d
Explanation: The reference cell contains a fixed quantity of gases. It contains CO, nitrogen and water vapour.

9. Hydrogen flame ionization detector can be used to detect only the peak of CO and not methane.
a) True
b) False
View Answer

Answer: b
Explanation: Hydrogen flame ionization detector can be used to detect the peak of both CO and methane. The accuracy is about 2%.
advertisement

10. CO analyzers based on infrared absorption would give greater sensitivity under which of the following conditions?
a) With larger cell path lengths
b) With smaller cell path lengths
c) By changing the gases in the reference cell
d) By increasing the sample flow rate
View Answer

Answer: a
Explanation: CO analyzers based on infrared absorption would give greater sensitivity with larger cell path lengths. If 1m path length is given, the range is from 0 to 50ppm.

Sanfoundry Global Education & Learning Series – Analytical Instrumentation.

To practice all areas of Analytical Instrumentation, here is complete set of 1000+ Multiple Choice Questions and Answers.

If you find a mistake in question / option / answer, kindly take a screenshot and email to [email protected]

advertisement
advertisement
Subscribe to our Newsletters (Subject-wise). Participate in the Sanfoundry Certification contest to get free Certificate of Merit. Join our social networks below and stay updated with latest contests, videos, internships and jobs!

Youtube | Telegram | LinkedIn | Instagram | Facebook | Twitter | Pinterest
Manish Bhojasia - Founder & CTO at Sanfoundry
Manish Bhojasia, a technology veteran with 20+ years @ Cisco & Wipro, is Founder and CTO at Sanfoundry. He lives in Bangalore, and focuses on development of Linux Kernel, SAN Technologies, Advanced C, Data Structures & Alogrithms. Stay connected with him at LinkedIn.

Subscribe to his free Masterclasses at Youtube & discussions at Telegram SanfoundryClasses.